M21-400 REV.1.1 DVPLUS DIGITAL VISCOMETER - OPERATIONAL MANUAL
DISPLAY READING WILL NOT STABILIZE
Special characteristic of sample fluid. There is no problem with the Viscometer.
Refer to Appendix C.
Check for erratic spindle rotation.
Verify power supply.
Contact AMETEK Brookfield or your local authorized dealer for repair (see Appendix M).
Bent spindle or spindle coupling.
Contact AMETEK Brookfield or your local authorized dealer for repair (see Appendix M).
Temperature fluctuation in sample fluid.
Use temperature bath for control.
Air Bubble under spindle
Viscometer not Level
Fluid Turbulence