EasyManua.ls Logo

Ametek JOFRA ATC-125 - MVI Circuit for ATC-250;320;650 A;B

Ametek JOFRA ATC-125
89 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
62 18-09-2007 122770 06
Adjustment – ATC-155 A/B
c
Turn the pre-set potentiometer R9 clockwise as far as
possible.
d
Switch the calibrator on using the power control switch.
e
Set the calibrator to –50°C.
f
Connect a voltmeter to CN8.
g
Adjust R9 until the voltage for CN8 reaches
a) 0.45V for PCB marked PCB 991025
b) 0.41V for PCB marked PCB 010125 or the latest PCB.
Adjustment – ATC-140/156/157 A/B
c
Turn the pre-set potentiometer R6/R16 counter clockwise as
far as possible.
d
Switch the calibrator on using the power control switch.
e
Set the calibrator to –50°C.
f
Connect a voltmeter to CN3.
g
Adjust R6 until the voltage for CN3 reaches 0.33V
h
Connect a voltmeter to CN6.
i
Adjust R16 until the voltage for CN6 reaches 0.33V
2.4.2 MVI Circuit for ATC-250/320/650 A/B
Necessary Equipment
Voltage stabilizer with output voltage 230V ±10V (or at 115V ±5V).
Set-up
PCB is adjusted when mounted in the calibrator.

Other manuals for Ametek JOFRA ATC-125