Versitron Operation Manual 59
Click the Apply tab will prompt the Capstan Wheel Adjustment screen, adjust the
capstan wheel to the optimum load position and when ready click Done.
The Add a New Reading screen is now ready for the first measurement to be
executed. Place the corresponding test block on the anvil and proceed to take the first
measurement by operating the arm lever or foot pedal, click OK if satisfy with the
reading or Cancel to retake the measurement.
When the first measurement is completed, the Indirect Calibration screen will appear
with the data grid listing the Reference value, the actual Reading and the Scale used.