ACA Current (Clamp-on 50Hz / 60Hz)
Range Accuracy
1) 2) 3)
40.00, 400.0, 600 A ±( 1.5% rdg + 8 digits)
Overload Protections: ACA Clamp-on jaws: 600 A rms continuous
ACD-14 Plus: Average Sensing
ACD-14 TRMS Plus: True RMS sensing - 10 % to 100 % of range
1) Max Induced error from adjacent current carrying conductor: 0.05 A
2) Specified accuracy is from 1% rdg to 100% rdg of range and for measurements
made at the jaw center. When the conductor is not positioned at the jaw center,
position errors introduced are: Add 2% rdg to specified accuracy for measurements
made BEYOND jaw marking lines (toward jaw opening)
3) Add 8 digits to specified accuracy @ reading < 10% rdg of range
Frequency
Function Sensitivity (Sine RMS) Range Accuracy
400.0 mVac 350mV 10 Hz ~ 2 kHz ±( 0.5% rdg + 4 digits)
4.000 Vac 1V 5 Hz ~ 5 kHz ±( 0.5% rdg + 4 digits)
4.000, 40.00 Vac 32V 5 Hz ~ 100 kHz ±( 0.5% rdg + 4 digits)
400.0 Vac 90V 5 Hz ~ 10 kHz ±( 0.5% rdg + 4 digits)
600 Vac 500V 5 Hz ~ 5 kHz ±( 0.5% rdg + 4 digits)
Display counts: 5000
Resolution: 0.001Hz
Transient protection : VAC input jacks : 6.5kV (1.2/50μs surge)
Ohms
Range Accuracy
400.0 ±( 0.8% rdg + 8 digits)
4.000, 40.00, 400.0 k ±( 0.6% rdg + 4 digits)
4.000 M ±( 1.0% rdg + 4 digits)
40.00 M ±( 2.0% rdg + 4 digits)
Open Circuit Voltage : 0.4 VDC typical
Transient protection : 6.5 kV (1.2/50μs surge)