EasyManua.ls Logo

Amprobe PM55A - General Specifications and Approvals; General Specs; Agency Approvals

Amprobe PM55A
80 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
11
Specifications
General Specifications
Display and Update Rate: 3-5/6 digits 6000 counts; Updates 5 per second
nominal
Operating Temperature: 0 °C - 40 °C
Relative Humidity: Maximum 80% R.H. up to 31 °C, decreasing linearly to
50% R.H. at 40 °C
Altitude: Operating below 2000 m
Storage Temperature: -20 °C ~ 60 °C, < 80% R.H. (with battery removed)
Temperature Coefficient: Nominal 0.15 x (specified accuracy)/ °C @
(0 °C ~ 18 °C or 28 °C ~ 40 °C), or otherwise specified
Sensing: Average sensing
Overload Protection: 600 V dc and V ac rms
Low Battery: Below approx. 2.4 V
Power Supply: 3 V standard button battery x 1 (IEC-CR2032; ANSI-NEDA-5004LC)
Power Consumption (typical): 6 mA for Voltage functions on Auto - V position and 2 mA for other
functions
APO Consumption (typical): 2.2 µA
APO Timing: Idle for 3 minutes
Dimension / Weight
L 113 mm x W 53 mm x H 10.2 mm / Approx. 78 gm
Special Features
AutoTect (Automatic V and selection) and VolTect Electric Field Detection
Agency Approvals
P >
Safety: Meets IEC61010-1, UL61010B-1, CAN/CSA-C22.2 No. 1010.1-92,
CAT II 600 V and CAT III 300V, Pollution Degree 2, Class 2
E.M.C. Meets EN61326 (1997, 1998/A1), EN61000- 4-2 (1995), and EN61000-4-3 (1996). This
product complies with requirements of the following European Community Directives: 89/ 336/ EEC
(Electromagnetic Compatibility) and 73/ 23/ EEC (Low Voltage) as amended by 93/ 68/ EEC (CE
Marking). However, electrical noise or intense electromagnetic fields in the vicinity of the equipment
may disturb the measurement circuit. Measuring instruments will also respond to unwanted signals
that may be present within the measurement circuit. Users should exercise care and take appropriate

Table of Contents