D625 Users Manual
29
15. Clean the D625 Meter flange face with mild
cleaner/detergent and rinse with water
16. Inspect the D625 Meter flange face for any
damage prior to reassembly.
17. Blow out the resonant tube orifice with
compressed air
18. Prepare the hardware for reinstallation
19. Coat each of the 5/16-18 x 1 ¼” Socket
Head CAP bolts with a small amount of anti-
seize compound
20. Reassemble the meter and install back into
the system.
21. Ensure all fittings are tightened with NO
LEAKS
22. Ensure the electrical connection between
the meter and density processor is securely
connected
23. Contact ANALYTICAL FLOW
TECHNOLOGIES with any questions or
concerns.
D625 METER MUST BE CLEANED AND RINSED PRIOR TO ANY FACTORY RETURN