D625 Users Manual
EXAMPLE: Field Calibration - K
0
Density Coefficient Offset Adjustment
Density Reading from D625
Meter
Density Error Offset (E
offset
)
E
offset
= D
actual
– D
meter
→ E
offset
= .99740 - .99865 → E
offset
= -
.00125
K
0(new)
= K
0(old)
+ E
offset
→ K
0(new)
= (-.433394018) + (-.00125)
Enter K
0(new)
into the Density processor or flow computer.
1
(Temperature of actual fluid density must be at the same temperature as the meter density
reading)
2
Calibrated Pycnometer, and/or a Hydrometer are examples of acceptable density test
equipment
FIELD CALIBRATION WORK TABLE
Field Calibration: K
0
Density Coefficient Offset Work Table
Density Reading from D625
Meter
Density Error Offset (E
offset
)
E
offset
= D
actual
– D
meter
→ E
offset
=___________-____________
K
0(new)
= K
0(old)
+ E
offset
→ K
0(new)
= ___________+___________
Enter K
0(new)
into the Density processor or flow computer.
1
(Temperature of actual fluid density must be at the same temperature as the meter density
reading)
2
Calibrated Pycnometer, and/or a Hydrometer are examples of acceptable density test
equipment