EasyManua.ls Logo

ANALYTICAL FLOW TECHNOLOGIES DENSITRAK D625 - Density Output - Field Calibration; When to Field Calibrate; Field Calibration Example

Default Icon
41 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
D625 Users Manual
30
DENSITY OUTPUT FIELD CALIBRATION
The ANALYTICAL FLOW TECHNOLOGIES D625 meter is calibrated at the factory with many various
fluids and various flow rates however, there may be occasions where field calibration (adjustment
of the K
0
density coefficient) is required. Many factors influence whether an adjustment of the K
0
density coefficient is required, such as:
Process pipeline flow rates
Noise or vibration in the process pipeline
Temperature differential
WHEN TO FIELD CALIBRATE
If the density meter is CONSISTENTLY displaying a density output reading that is above or below
the ACTUAL density of the fluid flowing through the pipe a simple adjustment of the K
0
density
coefficient can be performed so that the meter is calibrated to the actual fluid.
The K
0
density coefficient can be found on the Calibration Certificate that was included with your
density meter. The K
0
density coefficient can also be read (and displayed) by the included
Windows ® Software that shipped with your meter. The Windows® software is also required
when making adjustments to K
0
.
FIELD CALIBRATION EXAMPLE:
An example of a K
0
adjustment might be as follows:
Distilled Water at 68°F (and minimal pressure) flowing through the meter reads (outputs) 0.99865
gr/cm
3
on the meter, but we know (from tables, or hydrometer, pycnometer, etc) that distilled
water has a density of 0.99740 gr/cm
3
at 68°F. This means that the meter has an error offset of
+0.00125 gr/cm
3
at 68°F.
We can correct that error by simply adjusting the K
0
density coefficient by the amount of the error
offset, and then entering the new K
0
density coefficient into the density processor or flow
computer.
EQUATION 4: K
0
DENSITY OFFSET EQUATION

󰇛󰇜
󰇛󰇜


The table below guides you through the K
0
Coefficient adjustment for the above Example.
CONTACT ANALYTICAL FLOW TECHNOLOGIES ENGINEERS WITH ANY QUESTIONS CONCERNING
YOUR SPECIFIC APPLICATION.
When comparing density readings of any fluids the temperature (and pressure) of the
fluid must be consistent (the same) for each device measuring density.

Table of Contents