Service, maintenance, lamp replacement
SPECORD PLUS Edition 03/2013 41
Fig. 24 Protective cap of the desiccant cartridge
8.5 Replacing the fuses
WARNING! Electric shock!
Before replacing the fuses switch off the SPECORD PLUS from the mains switch and dis-
connect the mains cable from the device connection.
1. Switch off the device and disconnect the mains cable from the SPECORD PLUS.
2. Open the fuse holder (7 in Fig. 12) by pulling its cover.
3. Replace the faulty mains fuses.
Only use the following fuse: T 3.15 H/250 V, Type 19181
4. Close the fuse holder.
5. Connect the SPECORD PLUS to the mains cable.
6. Switch on the SPECORD PLUS.
The SPECORD PLUS can now be switched back on.