8-5. Function list
Class Item
Param-
eter
Description
0
Auto power-off disabled
poff
Auto power-off
1
Auto power-off enabled
Automatically
power off
0
Fast / sensitive
1
2
3
Cond
Response
4
Slow / stable
Software filtering
0
Stable when within ± 0.5d/0.5s
1
Stable when within ± 1d/0.5s
st-b
Stability band
width
2
Stable when within ± 2d/0.5s
Conditions to
turn on the
STABLE indicator
0
Disabled
trc
Zero tracking
1
Enabled
Tracking zero shift
0
Point (.)
pnt
Decimal point
1
Comma (,)
Decimal separator
0
Comparator disabled
1
Compares all data
2
Compares all stable data
3
Compares plus data > +4d
4
Compares stable plus data > +4d
5
Compares data > +4d or < -4d
Cp
Comparator mode
6
Compares stable data > +4d or < -4d
Conditions to
compare.
d = the minimum
display division
0
Command and stream modes
1
Command and PRINT key
2
Command, PRINT key and auto-print A
3
Command, PRINT key and auto-print B
prt
Data output
mode
4
Command mode only
Auto-print A:
+ data
Auto-print B:
+/- data
0
No pause (general equipment)
pUse
Data output pause
1
1.6 seconds (for AD-8121)
Interval between
continuous data
0
No output
1
AD-8121 format
(
*
)
info
GLP output
2
General format
GLP
output format
0
2400 bps
1
4800 bps
2
9600 bps
bps
Baud rate
3
1200 bps
0
7 bits, even parity
1
7 bits, odd parity
func
btpr
Data and parity
2
8 bits, non parity
Factory setting
(
*
)
When the AD-8121 format is selected, the interval between data is 1.6 seconds
regardless of the setting “pUse”.