DN:KD-7908-SMSY01 V2.0 Operation Guide
Template version:1
LCD Display shows
abnormal result
The cuff position was not correct
or it was not properly tightened
Apply the cuff correctly and
try again
Body posture was not correct
during testing
Review the ―BODY
POSTURE DURING
MEASUREMENT‖ sections
of the instructions and
re-test.
Speaking, arm or body
movement, angry, excited or
nervous during testing
Re-test when calm and
without speaking or moving
during the test
It is inappropriate for people
with serious arrhythmia to
use this blood pressure
monitor.
LCD shows low battery
symbol
Pressure system is unstable
before measurement
Don’t move and try again.
Fail to detect systolic pressure
Fail to detect diastolic pressure
Pneumatic system blocked or cuff
is too tight during inflation
Apply the cuff correctly and
try again
Pneumatic system leakage or
cuff is too loose during inflation
Cuff pressure above 300mmHg
Measure again after five
minutes. If the monitor is
still abnormal, please
contact the local distributor
or the factory.
More than 160 seconds with cuff
pressure above 15 mmHg
Device parameter checking error
Pressure sensor parameter error
No response when you
press button or load
battery.
Incorrect operation or strong
electromagnetic interference.
Take out batteries for five
minutes, and then reinstall
all batteries.