DN:KD-7908-SMSY01 V2.0 Operation Guide
Template version:1
9/15
Picture 7-6 Picture 7-7 Picture 7-8
d. When displaying the stored results, the monitor will turn off automatically after 1 minute of no operation.
You can also press the button ―START‖ to turn off the monitor manually.
8. DELETING MEASUREMENTS FROM THE MEMORY
When any result is displaying (exclude average value displaying), keep on pressing button
―MEM‖ for three seconds, all results in the current memory bank will be deleted after three
―beep‖. LCD will show picture 8, Press the button ―MEM‖ or ―START‖, the monitor will turn off.
Picture 8
9. ASSESSING HIGH BLOOD PRESSURE FOR ADULTS
The following guidelines for assessing high blood pressure (without regard to age or gender) have been
established by the World Health Organization (WHO). Please note that other factors (e.g. diabetes, obesity,
smoking, etc.) need to be taken into consideration. Consult with your physician for accurate assessment, and
never change your treatment by yourself.
Note: It is not intended to provide a
basis of any type of rush toward
emergency conditions/diagnosis based
on the color scheme and that the color
scheme is meant only to discriminate
between the different levels of blood
pressure.
10. TECHNICAL ALARM DESCRIPTION
The monitor will show 'SYS HI','SYS Lo' ,'DIA HI' or 'DIA Lo' as technical alarm on LCD with no delay if the
determined blood pressure (systolic or diastolic) is outside the rated range specified in part SPECIFICACIONS.
In this case, you should consult a physician or check if your operation violated the instructions.
The technical alarm condition (outside the rated range) is preset in the factory and cannot be adjusted or
inactivated. This alarm condition is assigned as low priority according to IEC 60601-1-8.
The technical alarm is non-latching and need no reset. The signal displayed on LCD will disappear automatically
after about 8 seconds.