10. Measurement range:
Cuff pressure: 0-300mmHg
Systolic: 60-260mmHg
Diastolic: 40-199mmHg
Pulse rate: 40-180 beats/minute
11. Accuracy:
Pressure: ±3mmHg
Pulse rate: ±5%
12. Environmental temperature for operation: 5℃~40℃(41℉~104℉)
13. Environmental humidity for operation: ≤90%RH
14. Environmental temperature for storage: -20℃~55℃(-4℉~131℉)
15. Environmental humidity for storage: ≤90%RH
16. Environmental pressure: 80kPa-105kPa
17. Battery life: Approx. 270 times
18. All components belonging to the pressure measuring system, including accessories: Pump, Valve, LCD, Cuff
and Sensor.
Note: These specifications are subject to change without notice.
NOTICE
1. Read all of the information in the operation guide and any other literature in the box before operating the unit.
2. Stay quiet, calm and rest for 5 minutes before blood pressure measurement.
3. The cuff should be placed at the same level as your heart.
4. During measurement, neither speak nor move your body and arm.
5. Measuring on same wrist for each measurement.
6. Please always relax at least 1 or 1.5 minutes between measurements to allow the blood circulation in your
arm to recover. Prolonged over-inflation (cuff pressure exceed 300 mmHg or maintained above15 mmHg for longer
than 3 minutes) of the bladder may cause ecchymoma of your arm.
7. Consult your physician if you have any doubt about below cases:
1) The application of the cuff over a wound or inflammation diseases;
2) The application of the cuff on any limb where intravascular access or therapy, or an arterio-venous (A-V)
shunt, is present;
3) The application of the cuff on the arm on the side of a mastectomy;
4) Simultaneously used with other monitoring medical equipments on the same limb;
5) Need to check the blood circulation of the user.
8. This blood pressure monitor is designed for adults and should never be used on infants or young children.
Consult your physician or other health care professionals before use on older children.
9. Do not use this unit in a moving vehicle, This may result in erroneous measurement.
10. Blood pressure measurements determined by this monitor are equivalent to those obtained by a trained
observer using the cuff/stethoscope auscultation method, within the limits prescribed by the American
National Standard, Electronic or automated sphygmomanometers.