DN:KD-7908-SMSY01 V2.0 Operation Guide
Template version:1
13/15
flicker emissions
IEC 61000-3-3
purposes.
Table 2
For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacturer’s declaration - electromagnetic immunity
The KD-7908 is intended for use in the electromagnetic environment specified below. The
customer or the user of the KD-7908 should assure that it is used in such an environment.
IMMUNITY test
IEC 60601test level
Compliance level
Electromagnetic
environment -
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be
wood, concrete
or ceramic tile. If
floors are
covered with
synthetic
material, the
relative humidity
should be at least
30 %.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency
magnetic fields
should be at
levels
characteristic of a
typical location in
a typical
commercial or
hospital
environment.
Table 3
For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacturer’s declaration - electromagnetic immunity
The KD-7908 is intended for use in the electromagnetic environment specified below. The
customer or the user of the KD-7908 should assure that it is used in such an environment.
IMMUNITY test
IEC 60601test level
Compliance
level
Electromagnetic environment
- guidance
Portable and mobile RF
communications equipment
should be used no closer to any
part of the KD-7908, including
cables, than the recommended
separation distance calculated
from the equation applicable to