Common Product Information
Document No. 992012. Issue 16 – 12.01.21
Page 23
The mattress is rated as medium to high risk and suitable for the
majority of patients up to 23 hours. It is important to remain
aware of individual patient needs, and standard nursing practices
must always apply for patients immobile or at high risk of
pressure sores.
When fitting a new mattress to the trolley the touch fastener on the patient
platform must also be replaced.
The mattress parts should be visually inspected for damage on a daily basis.
If the outer mattress fabric is torn, then fluids may penetrate and the
mattress should be replaced. Do not attempt to repair tears or splits with
self-adhesive tapes.
CAUTION: Ensure that the mattress is correctly orientated on the patient
platform with the touch fastener of the mattress aligning with the touch
fastener on the patient platform.
CAUTION: Ensure that the mattress is centrally positioned across the width of
the patient platform otherwise it may prevent the side rail from locking
when raised.
38. Cleaning and Disinfecting the Trolley
It is recommended that only CE marked cleaning products are used in the cleaning of
the trolley and the mattress. Cleaning and Disinfection should be carried out by hand
only.
Clean the trolley and mattress with warm water and neutral detergent and dry the
surfaces thoroughly using a soft cloth. Suitable disinfectants are: quaternary
ammonium compounds, isopropyl alcohol & chlorine bleach up to 1% (10,000 ppm).
Apply disinfectant by cloth, spray or disinfectant wipe. Following disinfection, wash off
all surfaces with clean warm water and dry thoroughly using a soft cloth. Clean all
touch fastener attachments periodically with a soft brush, neutral detergent and
suitable disinfectant as listed. The product will be adversely affected and its life
expectancy reduced if the above cautions are not observed.
CAUTION:
Do not steam clean or jet wash this device.
Do not soak or immerse this device.
Do not use concentrated bleaching disinfectant solutions, organic solvent or
abrasive powders in the cleaning or disinfection of this product.
Dilute all disinfectants in accordance with the manufacturer’s guidelines.
Disinfectant products are corrosive in nature; failure to properly wash and
dry the product surface could leave a corrosive residue which may cause
damage to the product.
Ensure the mattress is thoroughly dried before refitting.
39. Product Warranty
The product, when new, is guaranteed to be free from defects in materials and
workmanship and to perform in accordance with the manufacturer’s specification for a
period of one year from the date of invoice from Anetic Aid or their approved
distributor. Anetic Aid will repair or replace, at their discretion, any components found
to be defective or at variance with the manufacturer’s specification within this time at
no cost to the purchaser.
Protect your investment with a manufacturer backed service and
maintenance package; contact Anetic Aid for more details.
Warranty exclusions; the warranty does not provide cover for breakage or failure due to
tampering, misuse, neglect, accidents, modifications or shipping. The warranty is also
void if the product is not used in accordance with the manufacturer’s instructions or is