Chapter 4 OTDR Commands
4-18
4.4.14 SENSe:VOFFset Command
Syntax: SENSe:VOFFset<wsp><value>
Parameters: <value>
Floating point format
Range: Offset value can be set plus/minus current dynamic range. (–64.0 to 64.0)
Description: Set vertical offset for the displayed trace(s). Offset value units – DB.
Example: sens:voff –5.0
Response: None
Errors: (–224, "std_illegalParmValue, Invalid Parameter Value")
4.4.15 SENSe:VOFFset? Command
Syntax: SENSe:VOFFset?
Parameters: None
Description: Queries vertical offset for the displayed trace(s).
Example: sens:voff? --- 0<END>
Response: Current vertical offset value.
Errors: None
4.4.16 SENSe:ACURsor Command
Syntax: SENSe:ACURsor<wsp><value>
Parameters: <value>
Floating point format
Range: 0.0 – Current distance range.
Description: Set A cursor position. Cursor position units – KM.
Example: sens:acur 20.5
Response: None
Errors: (–224, "std_illegalParmValue, Invalid Parameter Value")
4.4.17 SENSe:ACURsor? Command
Syntax: SENSe:ACURsor?
Parameters: None.
Description: Queries current A cursor position.
Example: sens:acur? --- 20.5<END>
Response: Current A cursor position value.
Errors: None