EasyManua.ls Logo

Anritsu MT9083 Series - Page 71

Anritsu MT9083 Series
90 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4.4 SENSe Subsystem Commands
4-19
4
OTDR Commands
4.4.18 SENSe:BCURsor Command
Syntax: SENSe:BCURsor<wsp><value>
Parameters: <value>
Floating point format
Range: 0.0 – Current distance range.
Description: Set B cursor position. Cursor position units – Km.
Example: sens:bcur 20.5
Response: None
Errors: (–224, "std_illegalParmValue, Invalid Parameter Value")
4.4.19 SENSe:BCURsor? Command
Syntax: SENSe:BCURsor?
Parameters: None.
Description: Queries current B cursor position.
Example: sens:bcur? --- 20.5<END>
Response: Current B cursor position value in Km.
Errors: None
4.4.20 SENSe:LSALeft Command
Syntax: SENSe:LSALeft<wsp><start>,<stop>
Parameters: <start>
Floating point format
Range: –100.0 – 400.0.
<stop>
Floating point format
Range: –100.0 – 400.0.
Start value must be less or equal to stop value.
Description: Set start and stop for left LSA marker. Start and stop units – kilometers (KM).
Example: sens:lsal 0.0, 0.5
Response: None
Errors: (–224, "std_illegalParmValue, Invalid Parameter Value")
(–200, "std_execGen, LSA Inactive State")

Table of Contents