4.4 SENSe Subsystem Commands
4-19
4
OTDR Commands
4.4.18 SENSe:BCURsor Command
Syntax: SENSe:BCURsor<wsp><value>
Parameters: <value>
Floating point format
Range: 0.0 – Current distance range.
Description: Set B cursor position. Cursor position units – Km.
Example: sens:bcur 20.5
Response: None
Errors: (–224, "std_illegalParmValue, Invalid Parameter Value")
4.4.19 SENSe:BCURsor? Command
Syntax: SENSe:BCURsor?
Parameters: None.
Description: Queries current B cursor position.
Example: sens:bcur? --- 20.5<END>
Response: Current B cursor position value in Km.
Errors: None
4.4.20 SENSe:LSALeft Command
Syntax: SENSe:LSALeft<wsp><start>,<stop>
Parameters: <start>
Floating point format
Range: –100.0 – 400.0.
<stop>
Floating point format
Range: –100.0 – 400.0.
Start value must be less or equal to stop value.
Description: Set start and stop for left LSA marker. Start and stop units – kilometers (KM).
Example: sens:lsal 0.0, 0.5
Response: None
Errors: (–224, "std_illegalParmValue, Invalid Parameter Value")
(–200, "std_execGen, LSA Inactive State")