4.4 SENSe Subsystem Commands
4-21
4
OTDR Commands
4.4.24 SENSe:LOSS:MODE Command
Syntax: SENSe:LOSS:MODE<wsp><value>
Parameters: <value>
Integer format
Range: 0|1|2|3|4|5|6
0 = Splice Loss Mode
1 = Two-point Loss Mode
2 = Two-point LSA
3 = dB/km Loss Mode
4 = dB/km LSA Loss Mode
5 = Two-point Attenuation Corrected Loss Mode
6 = ORL Loss Mode
Description: Set current Loss Mode.
Example: sens:loss:mode 0
Response: None
Errors: (–224, "std_illegalParmValue, Invalid Parameter Value")
4.4.25 SENSe:LOSS:MODE? Command
Syntax: SENSe:LOSS:MODE?
Parameters: none
Description: Query current Loss Mode.
Example: sens:loss:mode? --- 0<END>
Response: Possible responses are:
0 = Splice Loss Mode
1 = Two-point Loss Mode
2 = Two-point LSA
3 = dB/km Loss Mode
4 = dB/km LSA Loss Mode
5 = Two-point Attenuation Corrected Loss Mode
6 = ORL Loss Mode
Errors: None