9.5. AIR-LIQUID SEPARATOR CANISTER BY CATTANI
At the beginning of each working day.
Insert inside filter (d) a defoaming tablet (v).
BIOLOGICAL HAZARD
Always wear gloves and goggles to prevent contact with
infected material when carrying out this operation.
At the end of each surgical procedure.
Perform an automatic flushing cycle or suck in about 1/2 litre of
sanitising solution with each one of the suction tubes used.
Sterilize the suction tube holder terminals in a steam autoclave up to
135°C following the instructions for use of the device.
At the end of each working day.
Suck in 1 litre of water with each suction tube, alternating water and air
(keep the suction tube alternately in and out of the water).
After rinsing with water, suck in about 1/2 litre of sanitising solution with
each one of the suction tubes used.
After flushing and disinfection, it is good practice to suck in air
only to dry the entire suction system (5 minutes).
Every 15 days.
Clean the separator container and probes with a soft non-abrasive
sponge and neutral detergent.
Clean the drain valve for the separator’s container with the brush
provided for this purpose.
Once a year.
By technician: check the siphons and drains, check all the internal tubes
and plastic and rubber parts subject to wear.
Before leaving the surgery for a few days.
Start suction and run it for 20-30 minutes without sucking in any liquids.
The suction unit will dry completely. As a result, salt caused by moisture
and basic substances will not form. Said salt may cause fan seizure and
motor blockage.
How to remove the separator’s container.
BIOLOGICAL HAZARD
Always wear gloves to prevent contact with infected
material when carrying out the following operation.
Dental units of A5 and A6 PLUS model
Move the dental chair fully up.
Open the water unit side cover.
Turn the electric box (b) after releasing the corresponding retainer.
Completely empty the separator container, pressing the proper timed
button (c) located on the cover.
If present, remove the valve (s) for centralised systems.
Turn and raise the container until it is detached from the drain pump (k).
Detach the container (d) from the cover (f) pulling up the two side rubber
bands (e).
After the cleaning operations, refit the container (d) after lubricating the
O-rings with S1-Protective Lubricant for O-Rings (CEFLA s.c.).