A3 PLUS - A5 - A6 PLUS - L6
SURGICAL SINGLE CART
EN
57
Maximum curing thickness.
The maximum curing thickness with every single cycle is 3 millimetres (refer to the instructions of the composite material used as well).
This thickness must not be exceeded as the layer may not be completely cured.
Cleaning.
The curing light may be a means of cross contamination between patients.
The most contaminated parts are the fibre optics and the anti-reflective cone.
Fibre optics and anti-reflective cone: sterilize in an autoclave using a sterilization temperature of at least 134°C for 3 minutes.
Before sterilizing them, make sure there are no residues of curing products: if necessary, clean with alcohol or a plastic spatula.
The fiber optic can undergo 500 autoclave cycles after which it tends to become opaque and therefore emit less light.
The eye protection must also be replaced after 500 cycles.
Contact the manufacturer to purchase original spare parts (fiber optic + eye protection: code 97660404).
Light handpiece: disinfect it on the outside with suitable products and, if necessary, use it by covering it with disposable film.
Use soft disposable paper towels to disinfect the handpiece. Do not use corrosive products.
The curing light handpiece is NOT suitable for autoclave.
The curing light handpiece is not protected against penetration of liquids therefore it CANNOT be soaked in a solution to be
sterilized.
The outside of the lamp should be disinfected with the fiber optic on.
Do not use any type of disinfectant on the exposed optical surface of the handpiece when the fiber is removed. The surface will
become irreparably opaque if it comes into contact with disinfectant.
Maintenance.
This equipment does not require any particular type of maintenance.
Only technicians authorised by the manufacturer can replace and/or repair the handpiece and dental unit.
The handpiece has been purposely constructed in a manner that requires specific tools to open it and therefore it cannot be removed by the user.
The warranty is automatically void if the handpiece is altered in any way.
Troubleshooting.
When the lamp is removed, the light does not come on (LED screen off).
Make sure the lamp is correctly attached to the power hose.
Carefully screw the ring nut, try to put the lamp back in and then take it out again.
If the problem persists, contact technical service department.
Less light emitted.
Make sure there are no residues of curing products on the end of the fibre optics: if necessary, wipe off with alcohol or a plastic spatula.
Make sure the fiber optic is not cracked or damaged in any way: replace it if it is so.
Contact the manufacturer to purchase original spare parts.
If the problem persists, contact technical service department.
When contacting the technical service department, please indicate:
The serial number of the handpiece engraved on the handpiece near the connector.
The type of defect.
If the handpiece must be sent back:
Please disinfect it first.
Ship it back in its original packing.
Please enclose a description of the fault with the shipping note.