10
4% HF (hydrofluoric acid) for 3-5 seconds, and wash it in distilled or deionized water, then submerge it in
the soaking solution for 24 hours to renew it.
4.5.5. Clean contaminated glass bulb and junction (please refer to Chart-6)
Chart – 6 Clean contaminated glass bulb and junction
Contamination
Cleaning Solutions
General tough contaminants
Apera Electrode Cleaning Solution (SKU: AI1166)
Organic lipid
Dilute detergent (weak alkaline)
Resin macromolecule
Dilute alcohol, acetone, ether
Proteinic sediment
Apera Electrode Cleaning Solution (SKU: AI1166)
Paints
Dilute bleacher, peroxide
Note: The electrode housing is polycarbonate. When use cleaning solutions, take cautions on
carbon tetrachloride, trichlorethylene, tetrahydrofuran, acetone, etc., which will dissolve the
housing and invalidate the electrode.
5
mV Value Measurement:
5.1 ORP Measurement
Press key, and switch the meter to mV measurement mode. Connect 301Pt-C ORP electrode (SKU:
AI1303, sold separately) and dip it in sample solution, stir the solution briefly and allow it to stay in the
solution until icon stays and then record the reading, which is ORP value. ORP is in short for Oxidation
Reduction Potential. The unit is mV.
5.2 Notes
5.2.1. ORP measurement does not require calibration. When the user is not sure about ORP electrode
quality or measuring value, use ORP standard solution to test mV value and see whether ORP electrode
or meter works properly.
5.2.2. Clean and Activate ORP Electrode
After the electrode has been used over long period of time, the platinum surface will get polluted, which
causes inaccurate measurement and slow response. Please refer to the following methods to clean and
activate ORP electrode:
(a) For inorganic pollutant, submerge the electrode in 0.1mol/L dilute hydrochloric acid for 30 minutes, then
wash it in distilled or deionized water, then submerge it in the soaking solution for 6 hours.
(b) For organic or lipid pollutant, clean the platinum surface with detergent, then wash it in distilled or
deionized water, then submerge it in the soaking solution for 6 hours.
(c) For heavily polluted platinum surface on which there is oxidation film, polish the platinum surface with