Revision 5 - 20 - November 2017
Disconnect black Safety Strap connecting
the chair to the tub.
Glide chair out of tub and fully onto the
carrier until the Chair Tab touches the stop
block and the Red Rail Caps are fully
behind the front edge of chair. Ensure that
the Primary Lock is engaged. Verify the
chair is locked to carrier by pushing the
chair back towards tub. If chair is secure, it
will not be able to move into the tub.
Unlock the casters on the carrier.