Page #25 background imageLoading...
Page #25 background image
Revision 5 - 25 - November 2017
Remedy® UV Infection Protection™ System (if applicable)
Your bathing system is equipped with Remedy® UV Infection Protection™ System, an
electronic germicidal ultraviolet (UV) light disinfecting system.
When you operate the bathing system, the bath water is continuously circulated through two
ultraviolet chambers, where the water passes over and around two UV lamps.
The UV light is monitored electronically, and the two green indicator lights show that the UV
light is sufficient to disinfect.
The Remedy® UV Infection Protection™ System is automatic; it turns on and off with the
whirlpool operation and has no separate controls.
Use only ultraviolet transmissible cleaners, disinfectants, soaps, and bath oils; such as Apollo's
Turbo Clean™ Pre-disinfectant Chelating Detergent, Cid-A-LII Quaternary Disinfectant,
Hygena™ Shampoo and Body Wash, and Therasol™ Natural Bath Oil. Use of other cleaners,
soaps, and bath oils is not recommended and could compromise the Remedy® UV Infection
Protection™ System.
Cleaning & Disinfecting
Your Advantage™ Bathing System arrived with an Advantage™ Intro Pack which includes
Turbo Clean™ Pre-disinfectant Chelating Detergent and Cid-A-L™ II Quaternary Disinfectant.
Turbo Clean™ is used after each bath to remove residue, bacteria and gross soils. Cid-A-L™ II
is used after each cleaning to effectively kill bath-borne bacteria, fungi and viruses.
Your Advantage™ Bathing System is calibrated to dilute Turbo Clean™ and Cid-A-L™ II at the
exact mixture stipulated by the manufacturer’s instructions. Use of other cleaners and
disinfectants is not recommended and could compromise the overall process.
With Remedy® UV Infection Protection™ System
Always follow the Cleaning Procedure after every bath.
It is not necessary to follow the Disinfecting Procedure after every bath in the same bathing
session.
For a Special Care Resident or in the event potentially infectious matter is released into the
tub (e.g. fecal matter, wound debridement, blood, mucus, etc.), consult infection control
management to determine if terminal disinfecting (Disinfecting Procedure) is required before
using again.
Always follow the Disinfecting Procedure after the last bath of the bathing session.
Without Remedy® UV Infection Protection™ System
Always follow the Cleaning and Disinfecting Procedure after every bath.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Apollo Advantage 6000 Series and is the answer not in the manual?

Apollo Advantage 6000 Series Specifications

General IconGeneral
BrandApollo
ModelAdvantage 6000 Series
CategoryBathtubs
LanguageEnglish

Related product manuals