CHAPTER 6
INSTRUMENTATION CALIBRATION
Entire Contents Copyright 2004 by Adaptive Power Systems, Inc. (APSI) • All Rights Reserved • No reproduction without written authorization from APSI.
APS-1000 Series 6 - 11 OM-001-01000-000-04.0
8. At the CURRENT Display, adjust the metered current using the ︿ or buttons
until the metered values match.
9. At the CURRENT Display, press the LOCK button to complete high-current
calibration.
10. Proceed to the next calibration function, unless this completes system calibration.
11. Note, to exit the Calibration Mode, your APS system must be restarted after you
have completed ALL calibration operations.