EasyManua.ls Logo

APS APS-1000 Series - Page 70

Default Icon
75 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CHAPTER 6
INSTRUMENTATION CALIBRATION
Entire Contents Copyright 2004 by Adaptive Power Systems, Inc. (APSI) • All Rights Reserved • No reproduction without written authorization from APSI.
APS-1000 Series 6 - 13 OM-001-01000-000-04.0
8. At the CURRENT Display, adjust the metered power using the ︿ or buttons
until the metered values match.
9. At the CURRENT Display, press the LOCK button to complete low-power
calibration.
10. Proceed to the next calibration function, unless this completes system calibration.
11. Note, to exit the Calibration Mode, your APS system must be restarted after you
have completed ALL calibration operations.

Table of Contents

Related product manuals