EasyManuals Logo

AquaLab 4TEV Series User Manual

Default Icon
122 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #113 background imageLoading...
Page #113 background image
16 APPENDIX B AquaLab
16 Appendix B
Temperature Correction
of Decagon’s Verification Standards
Table 5: Water Activity of Selected Salt Solutions
Temp. H
2
O
0.50 2.33 6.00 8.57 13.41
(
C)
mol/kg mol/kg mol/kg mol/kg mol/kg
KCL NaCL NaCL LiCl LiCl
15.0 1.000 0.984 0.923 0.761 0.492 0.238
20.0 1.000 0.984 0.922 0.760 0.496 0.245
25.0 1.000 0.984 0.920 0.760 0.500 0.250
30.0 1.000 0.984 0.920 0.760 0.504 0.255
35.0 1.000 0.984 0.920 0.760 0.508 0.261
40.0 1.000 0.984 0.921 0.760 0.512 0.266
50.0 1.000 0.984 0.894 0.740 0.517 0.275
Note: The AquaLab measures these verification standards to ±0.003
a
w
with the dew point sensor and ±0.015 a
w
with the capacitance
sensor.
108