AquaLab is the fastest and most accurate instrument for measuring
water activity, giving readings in five minutes or less. Its readings
are reliable, providing ±0.003 a
w
accuracy. The instrument is easy
to clean and checking calibration is simple.
Note: UL has listed this product to applicable UL Standards and
Requirements.
2.1 AquaLab 4 Instrument Specifications
Water Activity Range: 0.030 1.000 a
w
Water Activity Accuracy: ±0.003 (4TE Dew Point Mode) and ±0.015
(4TEV Capacitance Mode)
Water Activity Resolution: 0.0001
Read Time
1
: 5 min.
Sample Temperature Range: 15 to 50
C
Sample Temperature Accuracy: ±0.2
C
Sample Temperature Resolution: 0.01
C
Sample Temperature Adjustment Increments: 0.1
C
Sample Dish Capacity: 15 mL Full
Operating Environment: 4 to 50
C; 0 to 90% relative humidity
(non-condensing)
Case Dimensions: 26.7 x 17.8 x 12.7 cm
Weight: 3.1 kg
Case Material: POLYLAC PA-765 (ABS) with fire retardant
Display: 64 x 128 Graphical
1
On samples with no significant impedance to vapor loss.
4

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the AquaLab 4TEV Series and is the answer not in the manual?

AquaLab 4TEV Series Specifications

General IconGeneral
BrandAquaLab
Model4TEV Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals