AquaLab 17 APPENDIX C
tivity is obtained;
a
w
= exp(φγcM
w
) (3)
For example, equation 3 gives the a
w
in a 6 mol kg
1
NaCl solution,
(M
w
= 0.018 kg mol
1
, φ = 2, and γ = 1.271; from tables in Robinson
and Stokes, 1965) as
a
w
= exp(2 × 1.271 × 6 × 0.018) = 0.760 (4)
It is important to note that equation 3 has no explicit temperature
dependence. Available data on temperature dependence of γ indi-
cates variation is less than ±2% over the range 0 to 50
C for NaCl
(Lang, 1967) and KCl (Campbell and Gardner, 1971) and no other
terms have any temperature dependence.
A further advantage of unsaturated salts is that there is no solid
phase present to affect the water activity of the solution. Salt in sat-
urated solutions can exist in different states and result in uncertainty
in the water activity values.
Instructions for Using Decagon’s Verification Standards
Simply empty one vial of standard solution into a sample dish and
place the dish immediately into the AquaLab for measurement. Each
vial should fill a sample dish to just less than half full. Table 6 shows
the expected values.
Note: If you need to obtain a Safety Data Sheet (SDS) for any of
these standards, a printable version is available on our website at
http://www.decagon.com/education/safety-data-sheets/.
111

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the AquaLab 4TEV Series and is the answer not in the manual?

AquaLab 4TEV Series Specifications

General IconGeneral
BrandAquaLab
Model4TEV Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals