Using CPD Radar in Local View
When you activate CPD Radar Mode from the Operation Modes (by hitting key 1 or
highlighting with the cursor buttons and pressing <ENTER>), the default is Local view.
The following screen will appear:
The Quiver is constantly scanning for CPD
sources, and you will see slight changes on the
screen as the scanning occurs. To stop the Quiver
from scanning, you can press the Stop soft key which
will be renamed into Run key. Press the Run soft key
when you need it to resume scanning.
The Quiver display defaults to the medium Delay
Range of 0-25µs. If you want to change the delay
range, open menu item Delay scale by hitting key 1
and pressing <Enter>. The dialog box will open
where the following settings can be selected with ^
and v cursor keys.
Range can be
selected as 0-5µs, 0-25 µs, or a maximum of 0-50
µs.
Delay Range 0
50 µs and 0-5 µs
is shown to the
right.
It is
important to
note that the time delay span does not changing the instrument resolution of the CPD