PUMP CONFIGURATIONS
14.203.B_A707 en.SM.V4.3X ARGUS Medical AG 28.03.06 / PJ 9 / 48
2.6. Address list of the pump configuration (without ARGUS service)
The following list declares the possible configuration options which can be
performed on the pump keypad without using the PC.
All these options can also be configured by the PC-Software “ARGUS service”.
Address Default Function Unit Range
left display
Index
PC
right display
1 2 No Key ON/OFF only at stop valid - 0=No / 1=Yes
2 10 No Operation without drop detector -
0=No / 1=Yes
3 11 No Recall of the last used ml/h rate -
0=No / 1=Yes
4 19 No Buzzer at start -
0=No / 1=Yes
5 44 Yes Automatic pressure release -
0=No / 1=Yes
6 49 No Alarm acknowledge with key MODE -
0=No / 1=Yes
100 361 5 Key ON/OFF delay time • 0.1 s 0 - 31
101 362 8 Display brightness level 2 - 15
102 363 7 Buzzer volume level 5 - 10
103 316 9 Default pressure limit (IV-set 1) • 100 mbar 1 - 10
104 318 20 Drop rate window centre (IV-set 1) Drops/ml 10 - 65
200 317 250 Air bubble size (IV-set 1) µl 50 - 1000
399 - 707 Enter the calibration menu (volume) - 123
- 707 Enter the calibration menu (pressure) - 1234
Note!
The address does not correspond with the index used by the “ARGUS service” tool.