SAFETY STANDARD CHECK
14.203.B_A707 en.SM.V4.3X ARGUS Medical AG 28.03.06 / PJ 47 / 48
10. SAFETY STANDARD CHECK
Safety Standard Check (SSC)
ARGUS 707 V_en
Serial-no.: …………………………....………………..... Inventory-no.: ………………….…..…………….....
Hospital: …………………………………………………. Department: …………………………………….…..
Customer: ………………………………………………………………………………………………….
The SSC has to be performed at least every 24 months or after 10'000 hours of operation.
The check has to be done in accordance to the user- and service manuals.
1
Check if a software upgrade is required?
2
Visual check for damage, cleanness and com-
pleteness
- Housing, labels, accessories, connectors,
power cable, etc.
3
Test the function of the stop flow clamp - Proper movement of the clamp
- Display pump type and software release
- Display of 2, 4, 7, F., in numeric display
4
Keep "MODE" pressed while switching on the
pump
- Display of all operation- and alarm indicators
5
Connect/disconnect the pump to the mains - The indicator “external supply” turns on/off
6
Test the drop detector by simulating drops - Check the green “drop” indicator
- Door open the “door” indicator lights up
7
Test the door switch, open and close the door
- Door close the “door” indicator turns off
8
Open the door and remove any IV set - Indicator “air bubble” lights up
9
Install a water filled IV set, close the door - Indicator “air bubble” turns off
- The red indicator “drop” lights up (delayed)
10
Set rate to 333.3 ml/h, press "START",
disconnect the drop detector
- The acoustical alarm turns on
11
Press "MODE" - The acoustical alarm mutes
12
Check the external connector “nurse call” - Relay contact switches (see chapter 9)
Calibration of the pressure sensors
IV set type used:
13
Codan …………...……. Other ………………….
- See chapter 6.3. and 6.4. of the service
manual
14
Volume calibration - See chapter 6.5. of the service manual
15
Charge the battery while the pump is running 16
hours, at a rate of 30.0 ml/h
- The indicator “external supply” must light
Battery check at a rate of 30 ml/h.
16
Run the battery test until the pump switches off
automatically
- The green indicator "battery" lights up during
this test
- Check the pump internal time and data
17
Printout the pump history (refer to chapter 4.4.5.)
- Check the battery run time by checking the
latest history entries, > 4h 30min
Electrical test according to EN60601-1 - Visual check of mains connector
18
(all measurements made with a power cable 2.5m)
- Measurements attached
19
Charge the battery after this test!
The pump has passed the SSC and is safe for use
Date / Name: ……...…………………………………….. Signature: ………………………………............…
ARGUS Medical AG