PUMP CONFIGURATIONS
14.203.B_A707 en.SM.V4.3X ARGUS Medical AG 28.03.06 / PJ 10 / 48
2.7. Index list of the pump configuration (with ARGUS service)
Index PC Add. Default Function Unit Range
1 Yes Run indication by running decimal point - No / Yes
2 1 No Key ON/OFF only at stop valid - No / Yes
3 No Rate change allowed only at STOP - No / Yes
4 No Key STOP delayed # 361 - No / Yes
5 No Second entry of rate # 3=Yes - No / Yes
6 No Static alarm (staff alerting system) - No / Yes
7 No Display elapsed time in run mode # 8=No - No / Yes
8 No Display remaining time # 7=Yes - No / Yes
9
10 2 No Operation without drop detector - No / Yes
11 3 Yes Recall of ml/h rate at next power on - No / Yes
12 No Recall of ml total at next power on - No / Yes
13 No Recall of ml inf. at next power on - No / Yes
14 No SBS (step by step function) - No / Yes
15 No Display VTBI (volume to be infused) - No / Yes
16 No Display “Set-X” if only one set is enabled - No / Yes
17 Yes KVO (KOR) enabled # 60 - No / Yes
18 No Drop alarm only if bottle empty # 10=No, trA - No / Yes
19 4 No Buzzer at start - No / Yes
20 No Menu Clr (clear ml inf.) # 15=No - No / Yes
21
22 No Menu trA (transport) # 10, 18=No - No / Yes
23 Yes Menu PrL (pressure alarm limit) - No / Yes
24 Yes Menu CAP (battery capacity) - No / Yes
25 No Menu SEt Fill (prime IV set)
26 No Menu InF (ml inf. since last power on) - No / Yes
27 No Menu dLo (data lock) - No / Yes
28 No Menu Stb (stand by) - No / Yes
29 No Menu Med (medication name) - No / Yes
30 No Menu tM (timer alarm) - No / Yes
31
32 No Menu boL (bolus release) - No / Yes
33 No Menu boLr (bolus rate) # 32=Yes - No / Yes
34 No Menu tot (bolus total) # 32=Yes - No / Yes
35
36
37
38 No Automatic bolus application # 32, 34=Yes - No / Yes
39 No Neonatology mode accessible in PrL menu - No / Yes
40
41 No Clear ml/h after infusion completed - No / Yes
42 No Clear ml total after infusion completed # 41=Yes - No / Yes
43 No Air volume accumulated - No / Yes
44 5 Yes Automatic pressure release after occlusion - No / Yes
45 Yes Pressure display ON (LED bar graph - 20/40/60/80/100%) - No / Yes
46 No Pressure display with indicator # 45=Yes - No / Yes
47 No Stand by- and battery pre alarm low volume - No / Yes
48 No Flashing numeric display at alarm - No / Yes
49 6 No Alarm acknowledge only with key MODE - No / Yes