PUMP CONFIGURATIONS
14.203.B_A707 en.SM.V4.3X ARGUS Medical AG 28.03.06 / PJ 11 / 48
50 No Start with ≥ 1 bar pressure allowed - No / Yes
51 Yes Enable upstream occlusion detection - No / Yes
52
53
54
55 No Select RJ-45 connector for serial interface - No / Yes
56 Yes Medication display alternate # 29=Yes - No / Yes
60 No KVO only after infusion completed - No / Yes
65 No Clear and continue # 15=No - No / Yes
101 Yes Set 1 enabled - No / Yes
102 No Set 2 enabled - No / Yes
103 No Set 3 enabled - No / Yes
104 No Set 4 enabled - No / Yes
105 No Automatic return on default set 1 after power up # 101=Yes - No / Yes
201 1065 Correction at 1 ml/h ratio 900 - 1100
202 1054 Correction at 3 ml/h ratio 900 - 1100
203 1044 Correction at 5 ml/h ratio 900 - 1100
204 1035 Correction at 10 ml/h ratio 900 - 1100
205 1027 Correction at 25 ml/h ratio 900 - 1100
206 1020 Correction at 50 ml/h ratio 900 - 1100
207 1014 Correction at 75 ml/h ratio 900 - 1100
208 1008 Correction at 100 ml/h ratio 900 - 1100
209 999 Correction at 150 ml/h ratio 900 -1100
210 988 Correction at 200 ml/h ratio 900 - 1100
211 976 Correction at 250 ml/h ratio 900 - 1100
212 967 Correction at 300 ml/h ratio 900 - 1100
213 952 Correction at 400 ml/h ratio 900 - 1100
214 942 Correction at 500 ml/h ratio 900 - 1100
215 925 Correction at 750 ml/h ratio 900 - 1100
216 910 Correction at 999 ml/h ratio 900 - 1100
217 1000 Correction at 1200 ml/h
SET 1
Do not change
these values unless
explicitly trained
ratio 1000-1100
221 1065 Correction at 1 ml/h ratio 900 - 1100
222 1054 Correction at 3 ml/h ratio 900 - 1100
223 1044 Correction at 5 ml/h ratio 900 - 1100
224 1035 Correction at 10 ml/h ratio 900 - 1100
225 1027 Correction at 25 ml/h ratio 900 - 1100
226 1020 Correction at 50 ml/h ratio 900 - 1100
227 1014 Correction at 75 ml/h ratio 900 - 1100
228 1008 Correction at 100 ml/h ratio 900 - 1100
229 999 Correction at 150 ml/h ratio 900 -1100
230 988 Correction at 200 ml/h ratio 900 - 1100
231 976 Correction at 250 ml/h ratio 900 - 1100
232 967 Correction at 300 ml/h ratio 900 - 1100
233 952 Correction at 400 ml/h ratio 900 - 1100
234 942 Correction at 500 ml/h ratio 900 - 1100
235 925 Correction at 750 ml/h ratio 900 - 1100
236 910 Correction at 999 ml/h ratio 900 - 1100
237 1000 Correction at 1200 ml/h
SET 2
Do not change
these values unless
explicitly trained
ratio 1000-1100