PUMP CONFIGURATIONS
14.203.B_A707 en.SM.V4.3X ARGUS Medical AG 28.03.06 / PJ 12 / 48
238
239
240
241 1065 Correction at 1 ml/h ratio 900 - 1100
242 1054 Correction at 3 ml/h ratio 900 - 1100
243 1044 Correction at 5 ml/h ratio 900 - 1100
244 1035 Correction at 10 ml/h ratio 900 - 1100
245 1027 Correction at 25 ml/h ratio 900 - 1100
246 1020 Correction at 50 ml/h ratio 900 - 1100
247 1014 Correction at 75 ml/h ratio 900 - 1100
248 1008 Correction at 100 ml/h ratio 900 - 1100
249 999 Correction at 150 ml/h ratio 900 - 1100
250 988 Correction at 200 ml/h ratio 900 - 1100
251 976 Correction at 250 ml/h ratio 900 - 1100
252 967 Correction at 300 ml/h ratio 900 - 1100
253 952 Correction at 400 ml/h ratio 900 - 1100
254 942 Correction at 500 ml/h ratio 900 - 1100
255 925 Correction at 750 ml/h ratio 900 - 1100
256 910 Correction at 999 ml/h ratio 900 - 1100
257 1000 Correction at 1200 ml/h
SET 3
Do not change
these values unless
explicitly trained
ratio 1000-1100
258
259
260
261 1065 Correction at 1 ml/h ratio 900 - 1100
262 1054 Correction at 3 ml/h ratio 900 - 1100
263 1044 Correction at 5 ml/h ratio 900 - 1100
264 1035 Correction at 10 ml/h ratio 900 - 1100
265 1027 Correction at 25 ml/h ratio 900 - 1100
266 1020 Correction at 50 ml/h ratio 900 - 1100
267 1014 Correction at 75 ml/h ratio 900 - 1100
268 1008 Correction at 100 ml/h ratio 900 - 1100
269 999 Correction at 150 ml/h ratio 900 - 1100
270 988 Correction at 200 ml/h ratio 900 - 1100
271 976 Correction at 250 ml/h ratio 900 - 1100
272 967 Correction at 300 ml/h ratio 900 - 1100
273 952 Correction at 400 ml/h ratio 900 - 1100
274 942 Correction at 500 ml/h ratio 900 - 1100
275 925 Correction at 750 ml/h ratio 900 - 1100
276 910 Correction at 999 ml/h ratio 900 - 1100
277 1000 Correction at 1200 ml/h
SET 4
Do not change
these values unless
explicitly trained
ratio 1000-1100
310 999.9 Max. flow rate ml/h 1 - 999.9
311 999 Max. prime rate ml/h 1 - 999
312 1200 Max. bolus rate ml/h 1 - 1200
313 10 Max. bolus total ml 1 - 999
314 10 Upstream occlusion sensitivity factor 0 - 60
315 20 Air summation volume • 50 µl 2 - 40
316 103 7 Downstream default pressure limit PrL • 100 mbar 1 - 10
317 200 250 Air bubble size µl 50 - 1000
318 104 20 Drop rate window center drops 10 - 65
319
Set 1 parameters