PUMP CONFIGURATIONS
14.203.B_A707 en.SM.V4.3X ARGUS Medical AG 28.03.06 / PJ 13 / 48
320 999.9 Max. flow rate ml/h 1 - 999.9
321 999 Max. prime rate ml/h 1 - 999
322 1200 Max. bolus rate ml/h 1 - 1200
323 10 Max. bolus total ml 1 - 999
324 10 Upstream occlusion sensitivity factor 0 - 60
325 20 Air summation volume • 50 µl 2 - 40
326 7 Downstream default pressure limit PrL • 100 mbar 1 - 10
327 250 Air bubble size µl 50 - 1000
328 20 Drop rate window center drops 10 - 65
329
Set 2 parameters
330 999.9 Max. flow rate ml/h 1 - 999.9
331 999 Max. prime rate ml/h 1 - 999
332 1200 Max. bolus rate ml/h 1 - 1200
333 10 Max. bolus total ml 1 - 999
334 10 Upstream occlusion sensitivity factor 0 - 60
335 20 Air summation volume • 50 µl 2 - 40
336 7 Downstream default pressure limit PrL • 100 mbar 1 - 10
337 250 Air bubble size µl 50 - 1000
338 20 Drop rate window center drops 10 - 65
339
Set 3 parameters
340 999.9 Max. flow rate ml/h 1 - 999.9
341 999 Max. prime rate ml/h 1 - 999
342 1200 Max. bolus rate ml/h 1 - 1200
343 10 Max. bolus total ml 1 - 999
344 10 Upstream occlusion sensitivity factor 0 - 60
345 20 Air summation volume • 50 µl 2 - 40
346 7 Downstream default pressure limit PrL • 100 mbar 1 - 10
347 250 Air bubble size µl 50 - 1000
348 20 Drop rate window center drops 10 - 65
349
Set 4 parameters
350
351 4 Air volume summation time • 8 min 1 - 8
361 100 5 Key ON/OFF delay time # 4 • 1/10 s 0 - 31
362 101 8 Display brightness level 2 - 15
363 102 10 Buzzer alarm volume level 5 - 10
364
365 103 150 Pressure step for neonatology mode # 39=Yes mbar 60 - 250
366 1 Pressure display unit (mbar / mmHg / kPa / cmH2O / Psi) Enum 1 - 5
367
368 200 300 Battery capacity (discharge time) min 45 - 300
369 5 Automatic menu fall back delay time sec 5 - 30
390 0 Last service date in year year 0 - 99
391 0 Last service date in month month 0 - 12
392 0 Last service date in day day 0 - 31
393 0 Service interval in months (0 = disabled) month 0 - 24
394 0 Service interval in hours of operation (0 = disabled) hour 0 -10000
395
396 0 Pump serial number xxx 8 yyy xxx 8 yyy
397
398
- 399 707 Enter the calibration menu / clears protection key - 123/4