The calculated correction factor can be applied to IV-sets which are released in the
config. (index #101 - 104) only. Please be sure which IV-set you want to calibrate
before you press the “apply” button. Per ex-works settings, only IV-Set number 1 is
calibrated.
Important remark!
It is mandatory to perform also a pressure calibration according chapter 6.3. if a
new IV-set will be used.
We recommend to do a control measurement after changing the correction factor!
4.4. ARGUS service - Toolbox
With the “ARGUS service” PC-tool you can set the pump clock,
change PIN code, read and print out history, etc.
Select the next step by pressing one of the but-
tons (set clock, change PIN code, view history,
service interval or replicate).
4.4.1. ARGUS service - Toolbox - Pump clock
Use this feature to synchronize to
pump internal clock with your PC time.
Please note: The pump internal clock will set to
the central European time zone (Bern, CET,
GMT +1.00h) as ex-works settings, the pump
internal clock will not switch automatically between
summer and winter time. All history logs
(refer to chapter 4.4.5) will base on this time.
4.4.2. ARGUS service - Toolbox - PIN code
Use this feature to set the pump PIN code.