PUMP CONFIGURATIONS
14.203.B_A707 en.SM.V4.3X ARGUS Medical AG 28.03.06 / PJ 14 / 48
518 2
Permissions for serial communication
0 = none, 1 = query only, 2 = parametrising, 3 = remote control
Enum 0 - 3
522 No Allow change of ml total while infusing # 65=No - No / Yes
523 15 Alarm lamp brightness level 2 - 15
524 Display a department info text (after power up) char
1-16 ASCII
Using the “ARGUS service” tool, the complete and detailed pump configuration
can be done.
2.8. Special configuration options (with ARGUS service)
Enabling a new IV set
Follow the steps below to enable and configure a new IV set:
a) write “Yes” in the corresponding index 101, 102, 103, 104
b) make sure the set definitions (index 201 to 277) correspond to the IV sets used.
In case of doubt, contact ARGUS Medical AG or your local dealer.
c) configure the set parameters (index 310 to 349 and 351)
d) perform a complete set calibration as per chapters 6.3. to 6.5.