EasyManua.ls Logo

Argus 707 V - Page 32

Argus 707 V
49 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
MAINTENANCE
14.203.B_A707 en.SM.V4.3X ARGUS Medical AG 28.03.06 / PJ 31 / 48
2. Press “START/STOP” to
store the values for the up-
and downstream sensor.
3. Press “MODE”.
Display shows “CAL 1.0b”.
2. Select address 399
and press “MODE”.
4. Press “START/STOP”.
Display shows “Set - 1-”.
2. Press “START/STOP”.
Displa
y
shows “CAL. door”.
1. Press key "1" to select the set num-
ber
y
ou want to calibrate
(
1 to 4
)
.
6.3.2. Enter into the pressure calibration mode
6.3.3. Set selection before calibrating
6.3.4. Calibrating the offset of both sensors
1. Enter into the
configuration mode
(see chapter 2.3.).
123
4
.
39
9
3. Enter value 1234.
1. Remove the IV set and
open the door.
CAL. door
Before you select an
IV set, check chap. 2.8.