PUMP CONFIGURATIONS
14.164.A_A707en.SM.V1.01.doc ARGUS Medical AG 10.02.04 AS 4 / 38
2.2. Interrogation mode
With the interrogation mode you can read the present configuration of the pump without
the possibility to modify any configurations.
Flashing decimal points indicate which display is ready to accept an input by the keys 100,
10 & 1.
In the interrogation mode the left hand display shows the address and the right hand dis-
play shows the according value configured at this address. Please refer to chapter 2.6
where the meanings of the addresses are explained.
To modify any configuration data you have to go into the configuration mode.
1. To enter into the interrogation mode, switch the pump on
while keeping the keys “MODE” and START/STOP” pressed.
70
7
P
R
o
g
2. Press the
“START/
STOP”
key.
4. To quit the interrogation
mode press the
“ON/OFF” key at least 2
seconds.
. .
5
.
2
.
0
3. Enter the requested address (see chapter 2.6), e.g.
52.
The programmed value of the address 52 appears
in the right hand display.