PUMP CONFIGURATIONS
14.164.A_A707en.SM.V1.01.doc ARGUS Medical AG 10.02.04 AS 8 / 38
2.6. Address list of the pump configuration
Add.: Address Def.: Default value of the manufacturer
P: Protection grade (W = Write enabled)
Add. Def. P Function Range
0 0 W PIN code active 0=No; 1=Yes
1 1 W Run indication by running decimal point 0=No; 1=Yes
2 0 W Key [ON/OFF] only at STOP valid 0=No; 1=Yes
3 0 W Rate change only at STOP valid 0=No; 1=Yes
4 0 W Key [STOP] delayed (time at #361) 0=No; 1=Yes
5
6 0 W Static alarm (staff alerting system) 0=No; 1=Yes
7 0 W Display elapsed time in run mode (#8=0) 0=No; 1=Yes
8 0 W Select remaining time (#7=1) 0=No; 1=Yes
9
10 0 W Operation without drop detector 0=No; 1=Yes
11 1 W Recall "ml/h" (rate) at next power on 0=No; 1=Yes
12 0 W Recall "ml total" (end volume) at next power on 0=N o; 1=Yes
13 0 W Recall "ml inf." (volume inf.) at next power on 0=N o; 1=Yes
14 0 W SBS Step by step function 0=No; 1=Yes
15 0 W Display VTBI (Volume To Be Infused) 0=No; 1=Yes
16
17 1 W KVO (KOR, mode see (#60) 0=No; 1=Yes
18 0 W Drop alarm only if bottle empty
(#10=0, => 'trA' as default)
0=No; 1=Yes
19 1 W Buzzer at start 0=No; 1=Yes
20 0 W Menu 'CLr' (Clear "ml inf.") (#15=0) 0=No; 1=Yes
21
22 0 W Menu 'trA' (Transport)(#10=0) 0=No; 1=Yes
23 1 W Menu 'PrL' (Pressure alarm limit) 0=No; 1=Yes
24 1 W Menu 'CAP' (Battery Capacity) 0=No; 1=Yes
25 0 W Menu 'SEt Fill' 0=No; 1=Yes
26
27 0 W Menu 'dLo' (Data-lock) 0=No; 1=Yes
28 0 W Menu 'Stb' (Stand-by) 0=No; 1=Yes
29 0 W Menu 'MEd' (Medication-no.) 0=No; 1=Yes
30 0 W Menu 'tM ' (Timer alarm) 0=No; 1=Yes
32 0 W Menu 'boL ' (Release Bolus) 0=No; 1=Yes
33 0 W Menu 'boLr' (Bolus rate)(#32=1) 0=No; 1=Yes
34 0 W Menu 'tot ' (Bolus total)(#32=1) 0=No; 1=Yes
38 0 W BOLUS application automatic (#32=1,#34=1) 0=No; 1=Yes
40 0 W Demo mode (menus addr. 20..34 and 7 enabled) 0=No; 1=Yes
41 0 W Clear "ml/h" after infusion completed 0=No; 1=Yes
42 0 W Clear "ml total" after inf. completed (#41=1) 0=No; 1=Yes
43 0 W Air volume accumulated, see #350 and 351 0=No; 1=Yes
44 1 W Automatic pressure release after occlusion 0=No; 1=Yes
45 1 W Pressure display (bar graph) 0=No; 1=Yes
46 0 W Pressure display with indicator (#45=1) 0=No; 1=Yes
47 0 W Standby- and battery pre alarm low volume 0=No; 1=Yes
48 0 W Flashing numeric display at alarm 0=No; 1=Yes
49 0 W Alarm acknowledge only with key [MODE] 0=No; 1=Yes