REPLACEMENT OF PARTS
14.164.A_A707en.SM.V1.01.doc ARGUS Medical AG 10.02.04 AS 27 / 38
1. Remove the four screws on the back side.
6. REPLACEMENT OF PARTS
6.1. General
CAUTION!
The ARGUS 707 may only be used with accessories and spare parts which have been
approved by ARGUS Medical AG for safe technical use.
CAUTION!
If a door, a housing, a pressure sensor or a main board is replaced, a full calibration
(pressure sensors and volume calibration) is required.
6.2. Disassembling of the ARGUS 707
CAUTION!
Disconnect the mains cable from the power outlet before opening the housing!
Observe the antistatic protection rules when disassembling the ARGUS 707 (the use
of an antistatic table mat and a grounded clip are recommended).
6.2.1. Disassembly of the case