Caregiver Obligations
The Carendo shower and toilet chair is subject to wear and tear, and the following actions must be performed when
specified to make sure that the product remains within its original manufacturing specification.
PREVENTIVE MAINTENANCE SCHEDULE Carendo shower and toilet chair
WARNING
To avoid malfunction resulting in injury, make sure to conduct regular inspections and follow
the recommended maintenance schedule. In some cases due to heavy use of the product and
exposure to aggressive environment more frequent inspections should be carried out. Local
regulations and standards may be more stringent than the recommended maintenance
schedule.
NOTE
Product cannot be maintained and serviced while in use with the patient.
CAREGIVER OBLIGATIONS Action/
Check
Every
USE
Every
WEEK
Every
6 MONTHS
Every 2:nd
YEAR
Clean/Disinfect
X
Visually check safety belt
X
Visually check all exposed parts
X
Visually check mechanical attachments
X
Check hand control and cable
X
Check/clean castors
X
Check cushions/pillows
X
Visually check battery charger and battery
X
Lubricate locking pins
X
Replace the safety belt
X
WARNING
To avoid injury to both patient and caregiver, never modify the equipment or use
incompatible parts.