14
11. Technical Specification
Table 1:
Pump
Supply Voltage and Frequency: 230V / 50 Hz, 230V / 60 Hz, 220V / 60 Hz, 120V / 60 Hz, 100V / 50 Hz
100V / 60 Hz
Power Input: 35 VA MAX
Size: 133 x 152 x 275 mm
Weight: 2.7 kg
Case Material: Fire Retardant ABS Plastic
Plug Fuse Rating: 5A BS1362 (UK ONLY)
Pump Fuse Rating: T1AL 250 V
Degree of protection
against electric shock:
Class II, Double Insulated with Functional Earth
Type BF
Degree of protection
against liquid ingress: IPX0 - No protection
Mode of operation: Continuous
Cycle or Therapy Modes: 60 seconds total
12 seconds inflation
48 seconds deflation
Pressure Range: 30 - 60 mmHg (±10 mmHg)
ENVIRONMENTAL INFORMATION
Condition Temperature Range Relative Humidity Atmospheric Pressure
Operating +10 °C to +40 °C
(+50 °F to +104 °F)
30% to 75%
(non-condensing)
700 hPa to 1060 hPa
Storage and Transport
(Long Term)
+10 °C to +40 °C
(+50 °F to +104 °F)
20% to 95%
(non-condensing)
700 hPa to 1060 hPa
Storage and Transport
(Short Term)
-20 °C to +50 °C
(-4 °F to +122 °F)
20% to 95%
(non-condensing)
500 hPa to 1060 hPa
Products should be well packaged and stored in a well-ventilated, dry and non-corrosive environment.
NOTE
If this pump unit is subjected to extreme cold during transit or storage, initial
performance may be impaired. Connect to power supply and leave running for approx.
30 minutes before use.