18
Surge
IEC 61000-4-5
±0,5kV ±1kV; ±2 kV,
AC Mains, Line to
Ground
±0,5kV ±1kV, AC
Mains, Line to Line
±0,5kV ±1kV;
±2 kV, AC
Mains, Line
to Ground
±0,5kV ±1kV,
AC Mains,
Line to Line
Mains power quality should be that of
a typical commercial or hospital
environment.
Voltage dips,
short
interruptions
and voltage
variations
on power supply
input lines
IEC 61000-4-11
0 % UT; 0,5 cycle
At 0°, 45°, 90°,
135°, 180°, 225°,
270° and 315°
0 % UT; 1 cycle
and 70 % UT;
25/30 cycles
Single phase: at 0°
0 % UT; 250/300
cycle
0 % UT;
0,5 cycle
At 0°, 45°,
90°, 135°,
180°, 225°,
270° and
315°
0 % UT;
1 cycle and
70 % UT;
25/30 cycles
Single
phase: at 0°
0 % UT;
250/300
cycle
Mains power quality should be that of
a typical commercial or hospital
environment.
If the user of the pump requires
continued operation during power
mains interruptions, it is
recommended that the pump be
powered from an uninterruptible
power supply or a battery.
Note: UT is the AC mains voltage prior to application of the test level.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones
and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be
predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the measured field strength in the
location in which the product is used exceeds the applicable RF compliance level above, the product
should be observed to verify normal operation. If abnormal performance is observed, additional
measures may be necessary.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 1 V/m.