KEY: X = Action required on the part of the customer S = Action required on the part of a qualified service technician
6
Recharge battery for a minimum of 15 hours once every seven days. See
“Operating Instructions” for battery charging procedure.
X S X
Operate the lift through its full range from both the control panel and the
control handset in a normal and smooth manner.
X
With the jib in the fully lowered position, visually examine the exposed
lifting straps on each side of the mast for any cuts or fraying. If damaged
in any way, withdraw lifter from use immediately and have lift straps
replaced with new ones.
X
Open and close the chassis legs and check for full range of motion and
smooth movement.
X
Visually check the condition of the handset and its cable. If damaged,
withdraw from service immediately and replace with new cable and
handset assembly.
X
Visually check that all external fittings are secure, and all screws and nuts
are tight.
X
Make sure all castors rotate freely and that the two rear brakes lock.
Replace as required.
X
Make sure all castor-mounting pins are tight on the chassis and chassis
legs, and that the castor tread is not damaged. Replace as required.
X
Make sure all instruction labels are firmly attached and readable. Replace
as required. See “Operating Instructions” for location of labels.
X
Examine all exposed parts, especially where there is close contact with the
patient’s body. Make sure no cracks or sharp edges have developed that
could cause injuries or become unhygienic. Replace where necessary.
X
Test the “Automatic Cut-out” function.
See Note 1.
S
Action
Maxi Move Accessories
Slings Battery Charger
Before
Every Use
Every
Week
Every 12
Months
Every 2
Years
Before
Every Use
Every
Week
Every 12
Months
Every 2
Years
Before
Every Use
Every
Week
Every 12
Months
Every 2
Years
Preventive Maintenance Schedule
001.25065.EN rev. 9