Nimbus 4
Nimbus Professional
...with people in mind
Instructions for Use
649933EN Rev 3 • 09/2015

Other manuals for Arjohuntleigh NIMBUS 4

Questions and Answers

  • S
    stephanie75Jul 31, 2025
    What to do if Arjohuntleigh Health Care Products pump shows flashing and power fail alarm?
    • O
      osmithJul 31, 2025
      If the pump detects a mains power failure, re-apply mains power, or press and hold the Run/Standby button for 3 seconds to put the pump into Standby. If the power failure continues, switch to Transport mode and disconnect the tubeset. The mattress will stay inflated for up to 12 hours. If re-applying mains power doesn't solve the issue, contact a service engineer.
  • M
    monroecassieAug 4, 2025
    Why are no indicators illuminated on my Arjohuntleigh Health Care Products pump control panel?
    • D
      Dr. Dana FletcherAug 5, 2025
      There are several reasons why the indicators might not be lit. First, ensure the pump is receiving mains power and that the power cable is correctly fitted. If the pump is inflating the mattress, the indicators will extinguish when the operating pressure is reached. Check that the CPR control is fully closed, the tubeset connectors are securely connected and not damaged, and the Transport control is set to Normal. Also, ensure the tubeset is not kinked. If these steps don't resolve the issue, there might be a leak in the system, and you should contact a service engineer.
  • M
    moodykaylaAug 8, 2025
    Why won't the mattress cells inflate on my Arjohuntleigh Health Care Products?
    • M
      Mary MercerAug 8, 2025
      The mattress cells may not inflate if the vent valves are open, so ensure they are closed. Also, check that the CPR control is fully closed.
  • J
    James WadeAug 12, 2025
    How to troubleshoot no indicators on Arjohuntleigh NIMBUS 4 Medical Equipment pump control panel?
    • B
      Brandon WoodsAug 13, 2025
      If no indicators are illuminated on the Arjohuntleigh Medical Equipment pump control panel, here's what to check: * Ensure the pump has mains power and the cable is correctly fitted. * Verify the tubeset connectors are securely connected and undamaged. * Confirm the CPR control is fully closed and the Transport control is set to Normal. * Check the tubeset isn't kinked. If the indicators extinguish when the operating pressure is reached, this is normal. If problems persist, a leak in the system may exist, and you should call the service engineer.
  • D
    Donna SanchezAug 17, 2025
    What to do if Arjohuntleigh NIMBUS 4 Medical Equipment shows flashing and power fail alarm?
    • J
      Jacqueline MartinAug 17, 2025
      If your Arjohuntleigh Medical Equipment displays a flashing light and a power fail alarm, it means the pump has detected a mains power failure. You can try to re-apply mains power, or press and hold the Run/Standby button for 3 seconds to put the pump into Standby. If the power failure continues, switch to Transport mode and disconnect the tubeset. The mattress will stay inflated for up to 12 hours. If the pump still isn't working after restoring power, contact the service engineer.
  • S
    Sierra WardAug 22, 2025
    Why won't mattress cells inflate on Arjohuntleigh NIMBUS 4?
    • C
      Carol PhillipsAug 22, 2025
      If the mattress cells of your Arjohuntleigh Medical Equipment are not inflating, ensure that the vent valves are closed and the CPR control is fully closed. These steps should resolve the issue.
  • D
    dawn33Aug 27, 2025
    What to do if Arjohuntleigh NIMBUS 4 shows flashing and pump failure?
    • J
      Jason LewisAug 27, 2025
      If your Arjohuntleigh Medical Equipment is flashing and indicating a pump failure, stop using the pump and call the service engineer immediately. The pump likely needs a service.

Summary

General Safety

Introduction

Intended use

Specifies the product's primary application for pressure ulcer prevention and management up to 250kg.

Installation

Installing the Mattress

Instructions for attaching the mattress to the bed frame using fastener straps.

Installing the Pump

Guidance on positioning the pump (hanging or under-bed) and connecting to power.

Testing the Power Fail Alarm

Procedure to verify the functionality of the pump's internal rechargeable battery alarm.

Connecting the Tubeset

How to properly attach the tubeset connector to the mattress and pump.

Controls, Alarms and Indicators

Pump Controls

Explanation of the Run/Standby, Static Mode, and Alarm Mute buttons on the pump.

Pump Indicators

Description of the various lights (Run, Static, Alarm, Pressure, Wait) and their meanings.

Operation

Installing the System

Pre-operation checks including CPR unit, Transport control, and Vent Valves.

Inflating the Mattress

Instructions for connecting the pump and inflating the mattress to normal operating pressure.

Guidelines for Selecting Mattress Vent Valves to Open

Rules for deflating cells for permanent pressure relief or temporary nursing procedures.

CPR Control

Procedure for rapid deflation of the mattress for emergency CPR access.

Nimbus Professional Mattress: Patient Positioning Guide

Decontamination

Chemical Disinfection

Recommended disinfectants like chlorine-releasing agents or 70% alcohol.

Routine Maintenance

Troubleshooting

Arjohuntleigh NIMBUS 4 Specifications

General IconGeneral
ModelNIMBUS 4
ManufacturerArjohuntleigh
Weight Capacity250 kg (550 lbs)
Cycle Time10 minutes
TechnologyAlternating pressure
Therapy ModesAlternating
Mattress Height20 cm
Cover MaterialVapour permeable
Power Supply100-240V, 50Hz
Dimensions (L x W)200 cm x 86 cm