  • L
    Lisa ShannonAug 12, 2025
    What to do if the actuator 'stalls' during lift on Arjohuntleigh Maxi 500 Medical Equipment?
    • S
      Sheri BurnettAug 13, 2025
      If the actuator 'stalls' during lifting with your Arjohuntleigh Medical Equipment, try the following: 1. Replace the battery pack with a fully charged one. 2. Ensure that you are not exceeding the lifting capacity.
  • A
    Anna GrossAug 16, 2025
    What to do if the battery pack indicates it is fully charged but the Arjohuntleigh Maxi 500 Medical Equipment lift only works a few times?
    • A
      Amber GreeneAug 16, 2025
      If the battery pack shows fully charged when in the charger, but the Arjohuntleigh Medical Equipment lift only performs a few lifts, replace the internal batteries.
  • S
    susan07Aug 19, 2025
    What to do if the hand control does not respond on Arjohuntleigh Medical Equipment?
    • J
      Jessica PatelAug 20, 2025
      If the hand control of your Arjohuntleigh Medical Equipment is unresponsive, try the following: * Check the red emergency stop button on the control box. * Verify the connection of the hand control cable. * Ensure the battery pack is correctly installed and fully charged. * Test with a new, fully charged battery pack.
  • T
    Thomas SalazarAug 25, 2025
    What to do if the UP and DOWN buttons on the control box do not respond on Arjohuntleigh Maxi 500 Medical Equipment?
    • B
      Brian TaylorAug 25, 2025
      If the UP and DOWN buttons on the control box of your Arjohuntleigh Medical Equipment aren't working, make sure to: * Check the red emergency stop button on the control box. * Confirm the battery pack is correctly installed and fully charged. * Test with a new, fully charged battery pack.
  • M
    mitchellvictoriaAug 27, 2025
    What to do if the actuator does not respond on Arjohuntleigh Maxi 500?
    • S
      Scott PachecoAug 27, 2025
      If the actuator on your Arjohuntleigh Medical Equipment is not responding, try these steps: * Check the red emergency stop button on the control box. * Ensure the battery pack is correctly installed and fully charged. * Test with a new, fully charged battery pack. * Verify the hand control is connected. * Check the actuator is connected to the control box.
  • E
    Ethan HendersonAug 31, 2025
    Why is my Arjohuntleigh Maxi 500 control box beeping?
    • A
      April HigginsAug 31, 2025
      If you hear an audible 'beep' from the control box of your Arjohuntleigh Medical Equipment, replace the battery pack with a fully charged one.
  • M
    Michael BentonSep 3, 2025
    What to do if the 'Power on' light on the Arjohuntleigh Medical Equipment charger is not lit?
    • B
      Barry StoneSep 3, 2025
      If the 'Power on' light on your Arjohuntleigh Medical Equipment charger is not lit, check to ensure the charger is properly plugged into an electrical wall outlet.
  • G
    george23Sep 7, 2025
    What to do if the charger is plugged in, but the “Power on” light is not lit on Arjohuntleigh Maxi 500 Medical Equipment?
    • B
      Barbara ParkerSep 7, 2025
      If the charger is plugged in but the “Power on” light remains off on your Arjohuntleigh Medical Equipment, verify that there is power at the wall outlet.

Summary

Safety Instructions

Risk Assessment Checklist for Service Technician

Guidance for technicians to assess risks before performing any service or repair.

Preventive Maintenance Schedule

Technician Inspections

Visual guide to inspection points for annual maintenance checks.

Troubleshooting

Service Procedures

Checking the Battery and Charger

Steps to verify the condition and functionality of the battery and charger.

Full Feature Test

Tests all functions of the lift to ensure proper operation and safety.

Verifying the Anti-Crush System

Verifies the functionality of the anti-crush safety feature.

Replacing the Leg Actuator

Detailed procedure for replacing the leg actuator.

Replacing the Legs

Guide on how to replace the lift legs, including necessary steps and warnings.

Calibrating the Digital Scale

Procedure for calibrating the digital scale, including different processes.

Safe Working Load Test

Procedure for performing a safe working load test on the lift.

Error Messages

Electromagnetic Compatibility

Electromagnetic Compliance

Details compliance with regulatory standards for blocking EMI from external sources.

Electromagnetic Emissions

Information on RF emissions, harmonic emissions, and voltage fluctuations.

Electromagnetic Immunity

Specifies immunity levels to electrostatic discharge, electrical fast transient, surge, and power frequency magnetic fields.

Arjohuntleigh Maxi 500 Specifications

General IconGeneral
BrandArjohuntleigh
ModelMaxi 500
CategoryMedical Equipment
LanguageEnglish

Related product manuals