What to do if the lift arm “up/down” and “in/out” of the bathtub does not work in Arjohuntleigh Malibu?
- RrasmussenjoelAug 19, 2025
Check the hand control and cables, and replace them if necessary.
What to do if the lift arm “up/down” and “in/out” of the bathtub does not work in Arjohuntleigh Malibu?
Check the hand control and cables, and replace them if necessary.
What to do if the lift arm “in/out” of the bathtub does not work in Arjohuntleigh Malibu?
First, ensure the bathtub is in the lowest position; if not, lower it. Then, check the hand control and cables, replacing them if necessary.
What does it mean if the battery level LED is showing a steady yellow light and a beeping sound is heard when the lift is operated in Arjohuntleigh Malibu Medical Equipment?
This means that mains is off and the battery is low. If a bath has started, continue it. Wait for mains to return before starting a new bath. With mains on, check the battery level LED for a green light indication; when the battery level with green LED light is steady, the battery is charged.
What does it mean if the battery level LED is flashing yellow and a fast beeping sound is heard, and only the lift arm “out” of the bathtub can be operated on Arjohuntleigh Medical Equipment?
This means that the mains is off and the battery is almost empty. Finish the bath at once. Wait for the mains to return before starting a new bath. With the mains on, check the battery level LED for a green indication. When the battery level with green LED light is steady, the battery is charged.
What does it mean if HI/LO LED is flashing yellow when trying to lift arm “out” of bathtub on Arjohuntleigh Malibu Medical Equipment?
This means that the bathtub is not in the lowest position. Lower the bathtub to the lowest position.
What to do if it is only possible to lower the lift of Arjohuntleigh Medical Equipment?
The lifting device of this equipment is equipped with a safety nut. In this case, stop using the equipment and contact qualified personnel for repair.
What to do if nothing happens when pushing a button on Arjohuntleigh Medical Equipment?
Check that the emergency stop button is not pushed down; reset it until it pops up.
What to do if nothing happens when pushing a button on the hand control of Arjohuntleigh Malibu?
Replace the hand control.
Why is the battery level LED flashing green with mains on in Arjohuntleigh Medical Equipment?
The battery is charging. It is fully charged when the battery level with green LED light becomes steady. If the battery level LED never becomes steady, the battery needs to be replaced during the next service.
What to do if there is no water flow or bad water flow during bathtub filling and shower in Arjohuntleigh Malibu Medical Equipment?
Check the incoming water flow and check/clean the filters.
Critical warnings related to falling, entrapment, fire, electric shock, and cross-contamination.
Table outlining maintenance tasks and their frequencies.
Table listing common problems and their solutions.