15A2135L 14
Enter a number (not exceeding 20 digits) identifying the operator for
this calibration. This number appears on the Printout. For
alphabetic character entry, see Appendix E.
If the ID is entered incorrectly, press CLEAR and enter the correct
ID.
The next prompt will instruct to enter the Pipette ID:
Enter Pipette ID:
__
4.5 Entering Pipette ID Number
At the prompt:
Enter Pipette ID:
__
Key in the pipette ID (not exceeding 20 digits) and press ENTER. For
alphabetic character entry, see Appendix E. This ID will be printed on the
PCS Printout for pipette cross-referencing, and as a hard copy record.
The Instrument will print the information just entered and display:
*** Printing ***
Please Wait
Followed by:
Enter Pipette
Volume: μL
4.6 Entering Pipette Volume
Enter the volume (in microliters) to be dispensed, and press ENTER. The
CLEAR key can be used to correct an entry error prior to pressing the
ENTER key.
NOTE: This volume must be entered in microliters ranging from 0.1 μL through 5000 μL.