15A2135L 19
Following the completion and printing of the first calibration run, the
following prompt will be displayed:
Another Cal w/Vial?
(Yes/No)
Press YES to continue. The next prompt will instruct to enter an
Operator ID:
Enter Operator ID:
__
Enter a number (not exceeding 20 digits) identifying the operator for
this calibration. For alphabetic character entry, see Appendix E.
This number appears on the Printout.
If the ID is entered incorrectly, press CLEAR and enter the correct
ID.
The next prompt will instruct to enter the Pipette ID:
Enter Pipette ID:
__
Continue the calibration of the second pipette from Section 4.5,
“Entering Pipette ID Number.”
When the vial's maximum capacity is reached, the following prompt
will be displayed:
Result #XX XXX.X __
along with the alternating prompts:
Use New Vial? (Y/N)
and
End of Vial